DOE-STD-1054-93
3.4.4.3
Calibration of M&TE - On Site
All M&TE on-site calibration should be done using plant-approved
procedures. These procedures should include the following items:
identity of the equipment to be calibrated
calibration equipment and reference standards to be used
precautions and limitations
checks, tests, measurements, acceptance tolerances, accuracy,
precision, range, and specific parameters
step-by-step instructions regarding sequence, methods, data
sheet completion details, etc. (including as-found and as-left
conditions)
special instructions, such as environmental conditions, when
appropriate
reference standard used should have an accuracy at least four
times greater than the device under test. If this accuracy ratio
cannot be met, analysis of the errors should be estimated to
provide a valid uncertainty of the calibration process.
M&TE calibration data sheets should include provisions for recording
as-found data before any adjustments or repairs are made. Completed
data sheets should be reviewed, by the standards laboratory supervisor,
prior to release of the equipment to service to verify that acceptance
criteria are satisfied. When M&TE is found to be out of calibration
(gross error), an evaluation should be performed as described in
Section 3.4.5.  Completed data sheets should be filed with the
maintenance history.