to prevent accidental attraction of iron objects and prevent damage to the cryostat. This is
especially important for superconducting magnet systems.
3. Provide equipment supports and bracing adequate to withstand the forces generated during
fault conditions.
4. Appropriately ground electrical supply circuits and magnetic cores and provide adequate fault
protection.
5. Provide means for safely dissipating stored energy when excitation is interrupted or a fault
occurs.
6. Provide appropriate warning signs to prevent persons with pacemakers or similar devices
from entering areas with fields of greater than 0.001 Tesla.
7. Personnel exposure to magnetic fields of greater than 0.1 Tesla should be restricted.
8. When a magnet circuit includes switching devices that may not be able to interrupt the magnet
current and safely dissipate the stored energy, provide a dump resistor connected directly
across the magnet terminals that is sized to limit the voltage to a safe level during the discharge
and safely dissipate the stored energy.
10.10.2.3 OPERATION AND MAINTENANCE
Verify that any inductor is de-energized before disconnecting the leads or checking continuity or
10.10.3 ELECTRICAL CONDUCTORS AND CONNECTORS
The conductors and connectors covered here are only those used in unconventional applications.
10.10.3.1 HAZARDS
Examples of hazards are as follows:
1. Metallic cooling-water pipes that are also used as electrical conductors present shock hazards
(i.e., they may not be readily recognizable as electrical conductors).
2. Improper application or installation of connectors can result in overheating, arcing, and shock
hazards.
3. Hazardous induced voltages and arcing can result from inadequate separation between high-
and low-voltage cables.
4. Use of an improper cable for a given type of installation (routing) can result in a fire hazard.
10.10.3.2 DESIGN AND CONSTRUCTION
When working with special conductors and connectors for R&D applications, the following
guidelines shall be implemented for design and construction:
10-27