3.
Radiological instruments should undergo calibration prior to use following any preventive or corrective
maintenance or any adjustment that voids the previous calibration. A battery change is not normally considered
maintenance.
564 Calibration Facilities
1.
Radiological monitoring instrument inspections, calibrations, performance tests, calibration equipment selection,
and quality assurance should be performed in accordance with the recommendations of ANSI N323. Responsible
individuals should:
a.
Locate activities in a manner to control radiation exposure to operating personnel and to personnel in adjacent
areas
b.
Minimize sources of interference, such as backscatter and non-ionizing radiation, during the calibration of
instrumentation and correct for interferences as necessary
c.
Operate in accordance with the referenced standards
d.
Generate records of in accordance with the referenced standards.
2.
For organizations that do not possess or use their own calibration facilities, contracted calibration services should
be performed in accordance with the referenced standards.
5-21