Click here to make tpub.com your Home Page

Page Title: Equipment Maintenance and Calibration.
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home

   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 



DOE-STD-1112-98
5.6.3 Facilities and Equipment. The service laboratory shall have facilities and equipment adequate
to perform the types of service for which it requests accreditation. Facility services,
alternating current (AC) power, and environmental controls shall be adequate for the
measurement systems. These facilities should also be located such that background
fluctuations are minimized from work operations around the facility. Adequate backup
equipment or systems for key measurement systems should be available for use in case the
primary systems fail. Backup systems can either include alternate electronics and instrumenta-
tion or an ability to arrange for the services of another DOELAP accredited radiobioassay
service laboratory on an emergency basis.
5.6.4 Equipment Maintenance and Calibration. The laboratory shall maintain a preventive
maintenance program for equipment used in measurement systems or quality control checks.
When equipment used for measurements or quality control is subject to change due to use or
the passage of time, it shall be calibrated periodically. Calibration is performed by
measurements with a certified source, a derived source traceable to the National Institute of
Standards and Technology (NIST) or with a Transfer Reference Standard (TRS). For direct
radiobioassay, recalibrations shall be performed with the appropriate calibration and source
geometries or with derived source calibration phantoms, however calibration checks of
instrument performance can be performed without using DOELAP phantom geometries.
5.6.4.1 The proper performance of the measurement system shall be verified periodically with
appropriate calibration standards. Either the contractor or an external calibration service may
calibrate equipment or the measurement system. All calibrations and characterizations shall be
performed using certified NIST-traceable sources, derived NIST-traceable sources, or to
standards maintained by an equivalent foreign national standards authority when available.
The terms used in this technical standard for traceability conform to the American National
Standard ANSI N42.22-1995 Traceability of Radioactive Sources to the National Institute of
Standards and Technology (NIST) and Associated Instrument Quality Control.
5.6.4.2 The calibration standards and any influencing environmental conditions shall be
documented for all calibrations. Calibration records shall be made available for inspection
during the assessors' on-site visit. The traceability of the reference standards used shall be
verified during the inspection.
23


Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business