|
| DOE-STD-3028-2000
6.2.3
Requirements for Container Testing
1. Design Qualification Testing
a) The outer container shall remain leak-tight as defined by ANSI N14.5 after a free
drop of the Package (outer container, inner container, and simulated contents)
from a height consistent with the maximum credible accident for the storage
facility (for example, from a height of 35 feet for Oak Ridge Building 3019) onto a
flat, essentially unyielding, horizontal surface. The drop test shall follow the test
procedures specified in applicable portions of 49 CFR 178.603 and 49 CFR
178.601.
b) The inner container shall remain leak-tight as defined by ANSI N14.5 after a free
drop of the container (including simulated contents) from a four foot height onto a
flat, essentially unyielding, horizontal surface. The drop test also shall follow the
test procedures specified in applicable portions of 49 CFR 178.603 and 49 CFR
178.601.
c) The outer container shall remain leak-tight, as defined in ANSI N14.5, after a
hydrostatic proof-test to 1.5 times the Design Pressure. The test shall be
conducted as specified by 49 CFR 178.601.
2. Testing At Closure
Both the inner and outer containers shall be tested for leak-tightness, as defined in
ANSI N14.5, at their times of closure.
6.2.4
Other Requirements
1. Both the inner and outer containers shall have unique permanent identification
markings, such as by etching or engraving.
2. The outer container and the exterior surface of the inner container shall conform to
acceptable removable contamination requirements as defined by 10 CFR 835,
Appendix D for transuranics at the time of assembly and closure of the outer
container.
9
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |